4 - 57
Determining the Analog Output Abnormality Treatment Direction
Introduction
It is possible to determine the anal
g output direc
• The abnormality treatment direction is very important for securing
the safety of the overall control process. Determine the treatment
direction very carefully. Otherwise the equipment will be dam-
aged.
Caution
Default setting
"LOW"
Setting range
"LOW":
Minimizes the output.
"HIGH":
Maximizes the output.
"HOLD": Holds the pulse to the value obtaining just before the abnormality
occurred.
Open the analog output abnor-
mality treatment direction setup
screen by following the steps to
enter the Engineering Mode.
Touch the key.
Use the and keys to
determine the abnormality
treatment direction.
Touch the key to move the
cursor to the "#".
1
Screen
Procedure
Step
2
3
4
# ERROR OUT MODE
I. OUT LOW
# ERROR OUT MODE
I. OUT LOW
# ERROR OUT MODE
I. OUT HIGH
# ERROR OUT MODE
I. OUT HIGH