5 - 8
False Signal Input by Calibrator
Introduction
The electromagnetic flowmeter is provided with a dedicated calibrator, which
has the function of generating the same signal as the flow rate signal from the
detector. The converter function can be checked using this false signal.
When to use
When a problem occurs in the converter, use this method to judge whether
detector or the converter is responsible for the problem.
Preparation
Prepare the following equipment.
• Dedicated calibrator and dedicated cable
• Digital voltmeter
• Resistor (250
Ω
)
In addition, the value of the measuring span is needed as data to be input into
the dedicated calibrator.
How to check
Calibrate the electromagnetic flowmeter according to the calibrator’s operat-
ing manual. The code of the calibrator’s operating manual is CM2-MGZ100-
2001.