5 - 6
Checking the Contact Input/Output Loop
Continued
Setting range
(continued)
When 2-contact input is selected
Open the contact input loop
check screen by following the
steps to enter the Maintenance
Mode.
There is no cursor movement
on this display. The indication
o f O P E N o r C L O S E w i l l
change depending on the con-
nected contact input terminal
status.
1
Screen
Procedure
Step
2
When 2-contact output is selected
Open the contact output loop
check screen by following
steps to enter the Maintenance
Mode.
Touch the key to move the
cursor to the “CLOSE” or
“OPEN” that indicates the con-
tact output status.
Using the key, select the
contact output status to be
checked.
The two contact output termi-
nals close or open simulta-
neously. While this screen is
being displayed, the contact
status is output as indicated. It
is impossible to change the sta-
tus of only one contact output
terminal.
1
Screen
Procedure
Step
Touch the key to move the
cursor to “>.”
3
2
4
> ST. IN1 CLOSE
ST. IN2 CLOSE
> ST. IN1 CLOSE
ST. IN2 OPEN
> ST. OUT1 CLOSE
ST. OUT2 CLOSE
> ST. OUT1 CLOSE
ST. OUT2 CLOSE
> ST. OUT1 OPEN
ST. OUT2 OPEN
> ST. OUT1 OPEN
ST. OUT2 OPEN