5.1 Checking Functions
Checking the Input/Output Signal Loop
Introduction
The converter includes a constant current generator. Current from 0% to
115% of the flow rate signal can be set. Loop checking can be performed with
this function, as can loop checking of the pulse output and contact input/out-
put. This function can be executed in the Maintenance Mode.
Occasion
Use this function to check wiring and the operation of the equipment con-
nected to the converter in the measuring loop.
Type
These are the items used for the loop check.
•
Analog output
•
Pulse output
•
Contact input/output
•
Excitation current