4 - 49
Setting the Pulse Scale
Introduction
The flow rate per pulse can be set when a pulse output board has been selected
(additional specifications). This is used to set the pulse scale so that the fre-
quency on the upper right of the 16-digit display will not exceed 2,000 Hz.
When changing the pulse scale with a double range, use the wider range.
100.00 cm
3
/p
Default setting
Setting range
Within the range where the frequency will not exceed 2000 Hz
Unit: m
3
, l, cm
3
, t, kg, g
Open the pulse scale setup
screen by following the steps to
enter the Engineering Mode.
Use the key to move the
cursor to the desired numbers.
Use the and keys to
change the figures.
Touch the key to move the
cursor to the "#".
1
Screen
Procedure
Step
2
3
4
# PLUS 27.780 Hz
SCL 100.00 1 / p
# PLUS 27.780 Hz
SCL 100.00 1 / p
# PLUS 13.890 Hz
SCL 200.00 1 / p
# PLUS 13.890 Hz
SCL 200.00 1 / p