5 - 5
Checking the Contact Input/Output Loop
Introduction
Check the contact input terminal status on the LCD by turning ON/OFF the
contact input terminal of the electromagnetic flowmeter.
It is also possible to check the loop of the contact output signal by turning ON/
OFF the contact output terminal of the electromagnetic flowmeter.
The screen that will be displayed depends on what contact input/output func-
tions have been selected (additional specification).
Setting range
“CLOSE” “OPEN”
Open the contact input/output
loop check screen by following
the steps to enter the Mainte-
nance Mode.
Touch the key to move the
cursor to the “CLOSE” or
“OPEN” that indicate the sta-
tus of ST.OUT (abbreviation
of contact output).
1
Screen
Procedure
2
Using the key, select the
contact output status to be
checked.
While this screen is being dis-
played, the contact status is
output as indicated.
While this screen is being dis-
played, the indication of OPEN
or CLOSE will change accord-
ing to the contact input termi-
nal status.
Touch the key to move the
cursor to “>”.
3
4
5
Continued on next page
Default setting
Display current status
> ST. IN OPEN
ST. OUT CLOSE
> ST. IN OPEN
ST. OUT CLOSE
> ST. IN OPEN
ST. OUT OPEN
> ST. IN CLOSE
ST. OUT OPEN
> ST. IN CLOSE
ST. OUT OPEN