4 - 16
–
999999999 - 9999999999
1
Open the built-in flow counter
reset value setup screen by fol-
lowing the steps to enter the
Operator’s Mode.
2
Touch the key to move the
cursor to the desired numerals.
In this example, the cursor is
moved to the position of the
“1” by touching the key
several times. (You can also
hold down the key.)
Setting the Reset Value of the Built-in Flow Counter
Introduction
This sets and changes the integration starting value of the built-in flow
counter.
This function is used when a pulse output board has been selected (additional
specification).
Default setting
The reset value is set to “0000000000” at shipment.
Setting range
Screen
Procedure
Step
3
Touch the or key to set
the desired numbers.
In this example, the number
“1” is changed to “5” by four
touches of the key.
4
Touch the key to return
the cursor to the “*”.
*
CNT-RESET VALUE
0000010000
*
CNT-RESET VALUE
0000010000
*
CNT-RESET VALUE
0000050000
*
CNT-RESET VALUE
0000000000
Summary of Contents for MagneW3000 PLUS
Page 1: ...CM2 MGG210 2001 MagneW3000 PLUS Smart Electromagnetic Flowmeter Model MGG14C User s Manual ...
Page 13: ...xi MEMO ...
Page 25: ...1 12 ...
Page 41: ...2 16 MEMO ...
Page 53: ...3 12 MEMO ...
Page 114: ...4 61 MEMO ...