5 - 4
Setting range
Checking the Pulse Output
Introduction
The pulse output can be checked by using the electromagnetic flowmeter as a
constant current generator.
This screen will appear if pulse output has been selected (additional specifica-
tion).
Default setting
Display current output
000.0% - 115.0%
Open the pulse output check
screen by following the steps to
enter the Maintenance Mode.
Touch the key to move the
cursor to the numerical figure
to be checked. In this situation,
the pulse of the frequency cor-
responding to the indication
will be generated.
1
Screen
Procedure
Step
Using the and keys,
change the percentage to the
desired one. In the example at
right, a pulse of the frequency
corresponding to a flow rate of
100% will be generated.
Touch the key to move the
cursor to “>.” The pulse will be
generated according to the indi-
cation on the large 7-segment
flow rate display.
2
3
4
> OUTPUT CHECK
P. OUT 000 .0 %
> OUTPUT CHECK
P. OUT 000 .0 %
> OUTPUT CHECK
P. OUT 100 .0 %
> OUTPUT CHECK
P. OUT 100 .0 %