Appendix
Pges
6 - 10
ba43121e03
01/2011
Preparation
For use with the TresCon analyzer module, fill the prepared cleaning
solution into the empty container (1.5 L) and label the container
respectively.
1
Fill the well cleaned 10 L container with 5 L deionized water.
2
Put on protective goggles and gloves.
3
Carefully fill the portion of NaOH into the container.
4
Tightly close the container.
5
Carefully sway the container to dissolve the solids. When doing
so, the container noticeably warms up (particularly at the
bottom!).
6
Carefully fill the portion of Na
2
EDTA into the container.
7
Fill the container up to the shoulder ( 5 L ) with deionized water.
Slight overfilling of the container by approx. 1 cm is tolerable.
8
Tightly close the container.
9
Carefully rock the container to dissolve the solids.
10
Let the container cool down to body temperature.