Operation
Pges
2 - 14
ba43121e03
01/2011
2.5
View/change the AutoClean parameters
You can display or change the parameters of the automatic AutoClean
cleaning procedure.
Dependencies
z
An AutoCal procedure automatically follows an AutoClean
procedure.
z
The AutoClean time and AutoCal time are linked (see below).
Note
You can trigger an exceptional AutoClean procedure. For more
information, read section 3.1 S
TARTING
THE
A
UTO
C
LEAN
PROCEDURE
on
page 3-3 in this manual. The operating sequence is described in detail
there.
Changing the
AutoClean interval
The factory settings correspond to the values obtained from practical
experience in normal wastewater treatment plants. The AutoClean time
parameter is linked with the AutoCal time and can only be changed in
the
Change AutoCal Parameters
menu (see section 2.4, page 2-12).
Parameter
Values
AutoClean time
Display of the time; corresponds to
the AutoCal time (cannot be
changed here).
AutoClean interval
1, 2, 3, 6, 12, 24 h
1
Measuring mode
Select the TresCon Pges module with
68
.
2
Call up the
Parameters Pges
menu with
3
. The main menu
appears on the display.
3
Select the
AutoClean
menu item with
68
and confirm
with
3
. The prompt
Presentation of data/Change
appears
.
4
Select the
Change
mode with
68
and confirm with
3
.
5
Enter the PIN digit-by-digit with
68
, change the digit
with
57
and confirm with
3
.
The analyzer module switches to the STOP mode.