Maintenance
3 - 31
ba43121e03
01/2011
3.10.2 Replace lamp
This service step is required if the lamp is defective and must be
replaced. The halogen lamp has to be replaced completely, with the
holder. An AutoCal procedure (calibration) is triggered when the
subsequent measuring procedure starts.
Note
The lamp type and order number are given in section 6.1 A
CCESSORIES
AND
WEAR
PARTS
.
Procedure
1
Measuring mode
Select the TresCon Pges module with
68
.
2
Call up the
Parameters Pges
menu with
3
. The main menu
appears on the display.
3
Select the
Service
menu item with
68
and confirm with
3
. The service menu appears.
4
Select the
Test lamp/Replace lamp
menu item with
68
and confirm with
3
.
5
Undo the knurled-head screws on the front plate of the
thermoblock and remove the front plate.