Maintenance
Pges
3 - 34
ba43121e03
01/2011
3.11
Replacing the tubing system
The complete tubing system has to be replaced at regular intervals.
The following components are part of the tubing system:
z
Permeate tube
z
all tubing sections between consumable canister and thermoblock
z
all transition tubing pieces
z
all connection pieces (T-pieces and L-pieces)
z
all pump tubings
z
tubing at the thermoblock outlet
A year's supply of spare tubings and connection pieces is contained in
the scope of delivery of the analyzer module. For reordering, see
section 6.1 A
CCESSORIES
AND
WEAR
PARTS
When it is replaced depends on the sample preparation. The TresCon
has adjustable timers that indicate the maintenance activities required
when the relevant service intervals expire.
Reagent causes severe corrosion
Do not allow this substance to get into the hands of children. Wear
suitable protective gloves and goggles/face mask when working with
this substance.
If this substance gets into the eyes, rinse thoroughly with water and
consult a doctor.
Warning
Only trained personnel is allowed to work on the analyzing section.
Always follow the warning instructions on the chemical containers and
the safety datasheets!
Always use the proper safety equipment!
Note
You are informed when the tubing system must be replaced by a timely
message on the display (service intervals).