Pges
Operation
2 - 9
ba43121e03
01/2011
2.2.1
Manually starting an AutoCal calibration procedure
Outside the times defined in the AutoCal menu, you can start an
AutoCal calibration procedure as follows:
Note
If the time scheduled for the AutoClean (cleaning) procedure is
reached, an AutoClean is performed before the AutoCal procedure.
9
– After a waiting period, the
absorbance is measured
and stored.
Cal Standard B
9
– The photometer block is
emptied
AutoCal: Emptying the
system
1
Measuring mode
Select the TresCon Pges module with
68
.
2
Press the
1
key. The AutoCal sequence begins
immediately. First,
AutoCal: Empty system
appears on the
display.
Step
Instrument status/action
Display on the screen