Pges
Operation
2 - 21
ba43121e03
01/2011
2.7
View/change the Display parameters
You can display or change the setting of the dimension in this menu.
Conversion of the
unit
The reference unit is Pges [mg/L]. Based on this unit, the conversion to
µmol/L is done with the following correlation:
Note
The conversion of the dimension does not apply to the input of the
value of calibration standard B. This always remains unchanged and is
displayed in mg/L Pges.
Effects of the
conversion
The following table demonstrates what effects the adjusted citation
form has where:
Resolution
The resolution of the measured value depends on the selected dilution
(see section 2.11 V
IEW
/
CHANGE
THE
D
ILUTION
PARAMETERS
):
Inputs
1 mg/L Pges = 32.28514 µmol/L Pges
Measured value
Yes
Measuring range
Yes
Input and display of Cnominal
Yes
Concentration details for standard B
No
Relay functions
Yes (
General
Parameters
menu)
Recorder functions
Yes (
General
Parameters
menu)
Resolution / Dimension
Dilution
mg/L
µmol/L
None, or < 1+3
0.01
0.1
≥
1+3
0.1
1
Parameter
Values
Dimension
mg/L
µmol/L