Reagent A (B, C, D) empty
Reagent A
(B, C, D) empty
P
h2 > 50 cm
Display on the screen
Type of error
Pressure sensor
Replenish dist. H2O
H2O must be
filled
P1
-60 cm < h1 < 4 cm
H2O blocked
H2O blocked
P1
h1 < -60 cm
H2O empty
H2O empty
P1
h1 > 50 cm
Display on the screen
Type of error
Pressure sensor
Replenish standard B
Standard B
must be filled
P1
-60 cm < h1 <4 cm
Standard B blocked
Standard B
blocked
P1
h1 < -60 cm
Standard B empty
Standard B
empty
P1
h1 > 50 cm
Display on the screen
Type of error
Pressure sensor
Replenish NaOH
NaOH must be
filled
P1
-60 cm < h1 < 4 cm
NaOH blocked
NaOH blocked
P1
h1 < -60 cm
NaOH empty
NaOH empty
P1
h1 > 50 cm
Display on the screen
Type of error
Pressure sensor
Sample inlet blocked
Sample inlet
clogged
P1
h1 < -60 cm
Lack of sample
Lack of sample
P3
P3 < 50 mbar
Outlet blocked
Outlet clogged
P3 P3 > 200 mbar
Display on the screen
Type of error
Pressure sensor
Dist. H2O
Standard B
NaOH
Sample inlet