Maintenance
3 - 15
ba43121e03
01/2011
3.7
Empty reactor
This service routine empties the digestion reactor if, for example, a
measurement was stopped and there is still some sample in the
reactor.
Caution
The reactor is only emptied if the digestion temperature lies below
50 °C. If the digestion temperature lies above 50 °C, the reactor is
cooled until the temperature drops below 50 °C.
Duration
Approx. 1.5 minutes
Note
The pump empties the reactor via the thermoblock in the normal
sample sequence. You cannot quit the procedure.
Starting the procedure
1
Measuring mode
Select the TresCon Pges module with
68
.
2
Call up the
Parameters Pges
menu with
3
. The main menu
appears on the display.
3
Select the
Service
menu item with
68
and confirm with
3
. The service menu appears.
4
Select the
Empty reactor
menu item with
68
and
confirm with
3
. A prompt appears on the display.
5
Confirm the prompt with
3
after checking. After the last
confirmation, the procedure begins immediately and the
remaining duration appears on the display.