Pges
Appendix
6 - 5
ba43121e03
01/2011
TXRCALI
Reads the calibration data.
Instrument response:
If the last calibration was erroneous, the calibration data of the
erroneous calibration and that of the last valid calibration is displayed.
The data of the erroneous calibration is labeled with "err" next to the
time.
TXRCINT
Read the following settings:
z
AutoCal: Time, interval
z
AutoClean: Time, interval
Instrument response:
TXRCONZ
Read the current concentration.
Instrument response:
-----------------------------------------------------------
Calibration data
Pges module1
24.08.07 14:52:52
-----------------------------------------------------------
Calibration data of 23.08.07 10:59:06
Standard A : 0.00 mg/l PO4-P
ESTDA : 2.39 mE
Standard B : 1.60 mg/l PO4-P
ESTDB : 335.56 mE
a0 : -0.011
mg/l
a1 : 4.802 mg/l/E
delta : 1.000
-----------------------------------------------------------
TresCon>
--------------------------------------------
Pges module1
24.08.07 15:30:37
--------------------------------------------
AutoCal: 07:00, 24 h
AutoClean: 07:00, 24 h
--------------------------------------------
TresCon>