Pges
Overview
1 - 7
ba43121e03
01/2011
Installing the analyzer
module
Fig. 1-4
Remove the left cover plate
1
Switch off the TresCon at the power switch (on the mounting
stand).
2
Open the door of the TresCon.
3
At the mounting slot, undo the screws (B) of the left dummy
plate (see Fig. 1-4) and remove the dummy plate.
Keep the fixing screws. They will be required in step 5!
TresCon
TresCon N0x-N
Ser.-Nr. 00110020
WTW 82362 Weilheim Made in Germany
TresCon NOx-N
Reinigungslösung
Cleaning Solution
Ablauf
V2
Thermoblock
Probe
Sample
V1
H O dest.
H O dest.
2
2
blau
blue
blau
blue
rot
red
Pumpschlauch Standard/Probe
blau / blau / blau
Pump Tube Standard/Sample
blue / blue / blue
rot
red
P3
P1
10 fixing screws (B)