Pges
Maintenance
3 - 5
ba43121e03
01/2011
3.2
Fill system
This routine fills the wet chemical system with consumables. The
TresCon guides the user through the service. The steps and activities
required appear on the display. Only perform this service during initial
commissioning or when returning the instrument to service. The supply
and connection of the consumables to the analyzer module is
described in section 1.3.4 C
ONNECTING
THE
PERMEATE
AND
CONSUMABLES
.
Duration
Approx. 9 minutes
Note
You cannot quit the procedure. At the end of the filling procedure,
AutoClean (cleaning) and AutoCal (calibration) are triggered.
Starting the procedure
1
Check whether all the consumable containers are sufficiently
full and correctly connected (see section 1.3.4 C
ONNECTING
THE
PERMEATE
AND
CONSUMABLES
).
2
Measuring mode
Select the TresCon Pges module with
68
.
3
Call up the
Parameters Pges
menu with
3
. The main menu
appears on the display.
4
Select the
Service
menu item with
68
and confirm with
3
. The service menu appears.
5
Select the
Fill system
menu item with
68
and confirm
with
3
. A checklist appears on the display.
6
Confirm the prompts each time with
3
after checking. After
the last confirmation, the procedure begins immediately and
the remaining duration appears on the display.