Setting up the cool box
The shelf life of the reagents is limited. Thus, they must be kept cool
and dark even during operation. For this purpose, a cool box is set up
on the reagent tray.
Connecting the reagents
The reagents contain corrosive or irritating substances
The safety instructions that appear on the containers must be followed.
11
Place a cool box on the tray of reagents.
12
Connect the mains power supply of the cool box to a free
power socket. Observe the connection data.
13
Connect the plug of the cool box with the mains power supply.
14
Set the slide switch on the cool box to "Cold".
15
Press the "Power" key on the cool box (see the operating
manual of the cool box).
16
Place the full reagent containers in the cool box with reagent A
and B on the left and reagent C and D on the right.
17
The cool box has four lead-in ducts in the lid with short lengths
of tubing on the inside. Connect these lengths of tubing with
the screw caps (green) of the reagent containers. While doing
so, pay attention to the assignment according to the letters (A,
B, C; D).
18
Close the cool box.
19
Connect the four suction tubes (green) with the screw
couplings on the cool box. While doing so, pay attention to the
assignment according to Fig. 1-14.