Appendix
Pges
6 - 6
ba43121e03
01/2011
TXRGPAR
Read the instrument parameters.
Instrument response:
TXWCALINT12
Setting of the AutoCal interval, here: 12 hours.
Instrument response:
TXWCALTIMExx:yy
Set the AutoCal time (xx = hours; yy = minutes, 00 or 30 allowed only),
here: 03:00 hours.
Instrument response:
--------------------------------------------
Pges module1
24.08.07 15:32:01
--------------------------------------------
1.21 mg/L Pges
--------------------------------------------
TresCon>
-----------------------------------------------------------
Selected parameters
Pges module1
24.08.07 14:53:48
-----------------------------------------------------------
Display : Pges mg/l
Measuring range: 0.01 - 3.00 mg/l PO4-P
Standard B : 1.60
mg/l
AutoCal time: 07:00 h
AutoCal interval: 24 h
AutoClean time: 07:00 h
AutoClean interval: 24 h
Measuring interval : 10 min
AutoKorr : on
Manual correction : off
Attenuation: 2
AutoFlow
Fluid level measurement: on
Overpressure measurement: on
Dilution: None
-----------------------------------------------------------
TresCon>
--------------------------------------------
Pges module1
24.08.07 15:36:34
--------------------------------------------
AutoCal: 07:00, 12 h
AutoClean: 07:00, 24 h
--------------------------------------------
TresCon>