Pges
Maintenance
3 - 9
ba43121e03
01/2011
3.4
Replenish dist. H2O
This routine fills the wet chemical system with dist. H2O. The supply
and connection of the dist. H2O to the analyzer module is described in
section 1.3.4 C
ONNECTING
THE
PERMEATE
AND
CONSUMABLES
.
Duration
approx. 2 minutes
Note
You cannot quit the procedure.
1
Check whether the container of dist. H2O is sufficiently full and
correctly connected (see section 1.3.4 C
ONNECTING
THE
PERMEATE
AND
CONSUMABLES
).
2
Measuring mode
Select the TresCon Pges module with
68
.
3
Call up the
Parameters Pges
menu with
3
. The main menu
appears on the display.
4
Select the
Service
menu item with
68
and confirm with
3
. The service menu appears.
5
Select the
Replenish H2O dist
menu item with
68
and
confirm with
3
. A checklist appears on the display.
6
Confirm the prompts each time with
3
after checking. After
the last confirmation, the procedure begins immediately and
the remaining duration appears on the display.
Starting the procedure