4 Preparation
•
Battery not fully charged (< 95 % charge)
•
Battery temperature between 0 °C and 45 °C
The battery status LEDs flash green during the charging
process.
2. Once the status LEDs go permanently green: Release the
battery latch and take the battery out of the charging station.
Result
The battery is fully charged.
4.4.4 Changing the battery
Requirement
The replacement battery is fully charged.
1. If the device is not connected to line power: Switch off the
device
(see “6.2 Switching the device off”, page 119)
2. Take battery out of the battery compartment.
3. Slide the replacement battery into the battery compartment
until you hear it engage.
4. If necessary: Switch on the device
.
The symbol
appears in the display.
Result
The battery is changed.
Risk of injury from battery being incorrectly inserted!
An incorrectly inserted battery may lead to device failure and thus
injure the patient.
Always push the battery into the battery compartment until
you hear it engage and it is flush with the housing.