18 Technical data
18.13 Non-invasive blood pressure
measurement (NIBP monitoring)
18.14 Operation/data management
Specification
Non-invasive blood pressure
measurement
Manufacturer
SunTech Medical
Measuring method
Oscillometric, diastolic values correspond to phase 5
Korotkoff sounds
Measurement range (systolic):
• Adults
• Children
• Infants
40 mmHg to 260 mmHg
40 mmHg to 230 mmHg
40 mmHg to 130 mmHg
Measurement range (diastolic):
• Adults
• Children
• Infants
20 mmHg to 200 mmHg
20 mmHg to 160 mmHg
20 mmHg to 100 mmHg
Accuracy (during operation between 0 °C and 50 °C) ±3 mmHg
Specification
Operation/data management
Display
• Illuminated symbols
• Device status indicators
Audio output
• Voice prompts
• Alarm tones
• Audible signals
Session documentation
Automatic recording of measured values for ECG,
SpO
2
and NIBP and of event data
Data transmission via SD card
SD card with 32 GB
Data transmission via WiFi
• Data transfer types: 12-lead ECG, function check
results, session data, service data
• WiFi: 802.11a/b/g/n
• Security: WPA2-PSK
• Protocol: https
• Port: 443
Data transmission via Bluetooth
®
interface
Data transfer type: Session data
Data evaluation
Via DEFIview PC software