14 Maintenance
*Additional information for Germany:
In accordance with the Technical Safety Check
(“Sicherheitstechnische Kontrolle” in accordance with § 11 of the
German regulation MPBetreibV), we as the manufacturer
recommend that all accessories for use of the
MEDUCORE Standard
2
that are connected to it likewise be subject
to a safety check (STK) at the same interval as the device.
14.3 Sending in the device
1. Disassemble parts.
2. Clean and disinfect parts
(see “8 Hygienic reprocessing”,
3. Send parts to WEINMANN Emergency or to technicians
expressly authorized by WEINMANN Emergency.
Accessories*
Please follow the instructions for use supplied with the accessories.
Should the accessories not come with their own instructions for use, the
same intervals as for the device apply.
Part concerned
Interval
Maintenance by
Risk of infection from contaminated parts!
The device, accessories, and other parts may be contaminated,
and infect technicians with bacteria and viruses. Parts sent in
which are clearly contaminated will be disposed of at the cost of
the sender by WEINMANN Emergency or by technicians
authorized by WEINMANN Emergency.
Clean and disinfect parts before sending them in.
Do not send in parts which are potentially contaminated.