18 Technical data
18.16 Saving of session data
18.17 Electromagnetic compatibility (EMC)
Medical electrical equipment is subject to special precautions in
relation to electromagnetic compatibility (EMC). It must be
installed and put into operation in accordance with the EMC
information contained in the accompanying documentation.
In addition to the performance defined in the standard, this device
has the following essential performance:
•
Delivery of a current pulse for defibrillation and cardioversion
•
Distinction between cardiac rhythms to be defibrillated and not
to be defibrillated
•
Display of electrocardiogram (ECG)
•
Display of plethysmogram and hemoglobin oxygen saturation
(SpO
2
)
•
Display of systolic and diastolic arterial blood pressure (NIBP)
Storage medium
Memory hours assuming typical use
Internal device memory (100 MB)*
9 h 30 min
SD card WM 39510 (32 GB)*
1675 h
* Since different curve forms can be compressed to differing degrees, the actual number of storage hours
may deviate. The data are typical values.
Delay in treatment due to power supply network disruption!
Disruptions to the power supply network may cause visible
disruptions to the ECG. This may lead to incorrect measurement
results and alarms and in AED mode, lead to shock delivery at the
wrong time. This may delay therapy and injure the patient.
If there is major disruption to the power supply network, only
operate the device with a battery.