6 Operation
2. Analyze session data with the DEFIview PC software in
accordance with the DEFIview instructions for use.
Result
The session data have been analyzed using the DEFIview PC
software.
6.18 Reprocessing the device after use
1. Remove disposable articles from the patient and dispose of them:
•
Defibrillation electrodes
•
ECG electrodes
•
Disposable pulse oximetry sensor
•
Disposable NIBP cuff for newborns
2. Put new disposables with the device.
3. Hygienically reprocess the device, components, and accessories
(see “8 Hygienic reprocessing”, page 230)
When doing so, please note: It is recommended that hygienic
reprocessing is carried out immediately after use, as dried-on
residues are harder to remove.
4. Ensure that the device and printer batteries have a sufficiently
good battery status.
5. Insert new printer paper and have spare printer paper to hand
(see “4.7.3 Inserting printer paper”, page 98)
6. Stow components and accessories in the protective transport
bag
(see “4.2.1 Stowing components and accessories in the
protective transport bag”, page 76)
Alternatively:
Stow components and accessories on the portable unit
“4.2.2 Stowing components and accessories on the portable
unit”, page 83)
7. If required: Store the device, components, and accessories in
accordance with the conditions for storage
Result
The device is reprocessed following use.