258
EN MEDUCORE
Standard
2
WM
6840
1
0
4
/2
021
10 User menu
Alarm limit
settings Infant
Heart rate
q
35/min - 250/min,
in increments of 5
Here you can set the upper
( ) and lower ( ) limit
values as of which the device
is to trigger an alarm.
According to the values set,
the setting ranges of the
upper and lower limit values
depend on each other as
follows:
The setting range for the
upper alarm limit ends at
least one setting value above
the set value for the lower
alarm limit. The setting range
for the lower alarm limit ends
at least one setting value
below the set value for the
upper alarm limit.
200/min
Heart rate
r
30/min - 245/min,
in increments of 5
100/min
Pulse rate
q
35/min - 250/min,
in increments of 5
200/min
Pulse rate
r
30/min - 245/min,
in increments of 5
100/min
SpO
2
saturation
q
66 % - 100 %
95 %
SpO
2
saturation
r
65 % - 99 %
85 %
SpO
2
saturation
q
45 mmHg -
130 mmHg,
in increments of 5
100 mmHg
SpO
2
saturation
r
40 mmHg -
125 mmHg,
in increments of 5
50 mmHg
NIBP systolic
q
25 mmHg -
100 mmHg,
in increments of 5
70 mmHg
NIBP systolic
r
20 mmHg -
95 mmHg,
in increments of 5
30 mmHg
Automatic alarm
limits
-
10 %
20 %
30 %
Here you can set the
automatic alarm limits. The
device automatically specifies
the alarm limits for the
physiological alarms. The
deviation is 10 %, 20 % or
30 % from the measured
values at the time of
activation.
20 %
Parameter
Possible
values
Description
Factory
setting