6 Operation
9. If necessary: Press the
Back
function button.
Alternatively:
Press the NIBP button
for < 2 s.
The device exits NIBP function mode and switches to the set
mode.
10. Once the measurement is finished: Remove the NIBP cuff.
Result
An interval measurement has been carried out.
6.12.4 Carrying out venous stasis
With venous stasis (
IV
function button), the device inflates the NIBP
cuff and maintains this pressure for the time preset in the operator
menu (
Operator menu | NIBP settings | Venous stasis
duration
). The venous blood return flow is impeded and the user
can puncture one of the patient’s veins. Venous stasis can only be
carried out on the
Adult
patient group.
If you are the operator of the device and have access to the
operator menu, you can disable the
Venous stasis
function:
Operator menu | System settings | Disable functions | Venous
stasis
(see “13.12 System settings”, page 342)
Requirement
•
The device is switched on
(see “6.1 Switching on the device”,
.
•
The
Adult
patient group has been selected.
•
A mode is set.
•
A non-invasive blood pressure measurement (NIBP
measurement) is prepared
(see “6.12.1 Preparing non-invasive
blood pressure measurement (NIBP measurement)”,
page 182)
.
1. Press the NIBP button
for < 2 s.
The device switches to NIBP function mode.
2. Press the
IV
function button.
The NIBP cuff is inflated to the pressure set in the operator
menu. The timer displaying the duration of venous stasis
counts down. Throughout venous stasis duration, pressure in
the NIBP cuff is maintained.