6 Operation
14. If desired (only with E-mail delivery 12-lead ECG option): Press
the
Send
function button.
15. Select the recipient's name and send the 12-lead ECG to a
predefined recipient by e-mail
delivery 12-lead ECG (only with E-mail delivery 12-lead ECG
option)”, page 299)
Following successful delivery, the symbol
appears in the
display. If delivery is unsuccessful, the symbol
appears in
the display and an alarm is displayed
.
Delayed therapy due to severely delayed reception of the
12-lead ECG!
If the infrastructure of the monitoring station/hospital is not
configured by the operator to receive 12-lead ECGs, receipt of the
12-lead ECG may be severely delayed. This may lead to delayed
therapy and injure the patient.
Configure e-mail delivery 12-lead ECG correctly and check it
.
If transmission time is unexpectedly high: Contact the system
administrator of the e-mail recipient.