5 Function check
1. Select the
Upload
function button.
All configured networks are displayed.
2. Select a network within range.
3. Confirm with the
OK
function button.
The device connects to the selected network and uploads the
result of the function check to WEINMANN Connect. When
upload is successful, the symbol
appears in the display. If
upload is unsuccessful, the symbol
appears in the display
and an alarm is displayed
4. Confirm with the
OK
function button.
Result
The function check has been uploaded to WEINMANN Connect.
5.4 Checking ECG cables
In addition to visual inspection of the ECG cables
Carrying out a function check”, page 104)
,
WEINMANN Emergency recommends checking the function of
ECG cables at regular intervals.