10 User menu
Parameter
Possible
values
Description
Factory
setting
Alarm limit
settings Adult
Heart rate
q
35/min - 250/min,
in increments of 5
Here you can set the upper
( ) and lower ( ) limit
values as of which the device
is to trigger an alarm.
According to the values set,
the setting ranges of the
upper and lower limit values
depend on each other as
follows:
The setting range for the
upper alarm limit ends at
least one setting value above
the set value for the lower
alarm limit. The setting range
for the lower alarm limit ends
at least one setting value
below the set value for the
upper alarm limit.
120/min
Heart rate
r
30/min - 245/min,
in increments of 5
50/min
Pulse rate
q
35/min - 250/min,
in increments of 5
120/min
Pulse rate
r
30/min - 245/min,
in increments of 5
50/min
SpO
2
saturation
q
66 % - 100 %
100 %
SpO
2
saturation
r
65 % - 99 %
85 %
NIBP systolic
q
45 mmHg -
260 mmHg,
in increments of 5
220 mmHg
NIBP systolic
r
40 mmHg -
255 mmHg,
in increments of 5
75 mmHg
NIBP diastolic
q
25 mmHg -
200 mmHg,
in increments of 5
110 mmHg
NIBP diastolic
r
20 mmHg -
195 mmHg,
in increments of 5
35 mmHg