6 Operation
6.16.4 Printing a defibrillation report
Requirement
•
A printer has been connected to the device
Connecting the printer to the device”, page 196)
•
The
Defibrillation report
menu item is activated in the
(see “13.9 Printer settings (only with Printing
1. Carry out defibrillation or cardioversion.
On every shock delivery, the printer prints a defibrillation report
(3 s before and 5 s after the shock delivery, shown in abstract
form here):
Result
A defibrillation report has been printed.
6.16.5 Printing a function check report
Requirement
A printer has been connected to the device
Connecting the printer to the device”, page 196)
.
1. Carry out a function check