14 Maintenance
14 Maintenance
14.1 General instructions
Maintenance, Technical Safety Check (in Germany only), and
measures such as inspections and repairs must only be carried out
by the manufacturer or by a technician expressly authorized by it.
14.2 Intervals
Risk of injury from impaired device functions or defects, lack
of testing and excessive service life!
If the device is not correctly tested, or is operated for too long,
impaired device functions, deviations or defects may occur. The
device cannot be used correctly, and may injure the patient and
the user.
Observe the intervals for maintenance, for Technical Safety
Check (“Sicherheitstechnische Kontrolle” in accordance with
§ 11 of the German regulation MPBetreibV), and for
metrological check.
Do not operate the device for longer than 10 years.
Part concerned
Interval
Maintenance by
Device
Maintenance-free
Technical Safety Check every year Manufacturer or a technician
expressly authorized by
manufacturer
Metrological check every two years
Battery
Maintenance-free
When stored in the device: Charge every 3 months.
When not stored in the device: Charge batteries with serial numbers
< 20,000 every 5 months and batteries with serial numbers
≥
20,000
every 9 months.
Recommendation: Replace battery after 2 years.
Charging station
Maintenance-free
Recommendation: Replace
charging station after 12 years