11 Alarms and faults
ECG module defective
Internal ECG module defective
Have the device repaired.
E-mail delivery error
(Alarm deactivated after 60 s)
Server not accessible
Check connection to network
.
Connection between device and
network interrupted
Check preset e-mail recipient
.
Check connection to network
. If necessary: Select a
different network.
Check date/time
Excessive NIBP cuff pressure
(Alarm deactivated after 10 s)
Pressure in the NIBP cuff increased
to such a degree as a result of
external influences that the safety
valve opened
Check which influences resulted in
the pressure increase and
eliminate them before the next
measurement.
Heart rate
q
Measured heart rate is above the
set upper alarm limit
If medically indicated, carry out
suitable treatment.
Adapt alarm limits
Heart rate > 250/min
Measured heart rate is above 250/
min (device shows --- as heart
rate)
If medically indicated, carry out
suitable treatment.
Heart rate
r
Measured heart rate is below the
set lower alarm limit
If medically indicated, carry out
suitable treatment.
Adapt alarm limits
Heart rate < 30/min
Measured heart rate is below 30/
min (device shows --- as heart
rate)
If medically indicated, carry out
suitable treatment.
Insert printer paper
Printer paper completely used up
Change printer paper
.
No printer paper inserted
Change printer paper
.
Memory error
Internal memory module defective Have the device repaired.
NIBP diastolic
q
(Alarm deactivated after 10 s)
Measured diastolic blood pressure
is above the set upper alarm limit
If medically indicated, carry out
suitable treatment.
Adapt alarm limits
Alarm
Cause
Remedy