6 Operation
2. If the patient group is incorrect: Select another patient group
(see “6.4 Selecting patient group”, page 120)
The NIBP module is configured accordingly in the device with
the selected patient group.
3. Press the
Interval
function button.
4. Set the interval duration using the navigation knob.
5. Press the
Start
function button.
Alternatively:
Press the NIBP button
for > 2 s.
An NIBP measurement starts. After the end of the
measurement, the device displays the systolic and diastolic
arterial blood pressure, the timer counts down, and when it has
finished the next measurement starts automatically.
6. Evaluate the NIBP measurement result.
7. If necessary: Cancel the NIBP measurement with the
Stop
function button.
The device releases the pressure from the NIBP cuff.
8. If necessary: Set alarm limits in the user menu
.
Alternatively:
Set automatic alarm limits in the application menu
Risk of injury due to incorrectly selected patient group!
The device only delivers correct measured values if the appropriate
patient group is selected. An incorrect patient group may lead to
incorrect measurements and injure the patient.
Adapt the patient group to the patient.
If the patient group is incorrect: Change the patient group in
the application menu.