EN MEDUCORE
Standard
2
WM
6840
1
0
4
/2
021
6 Operation
19. Press and hold both
(
SHOCK
) buttons on the paddles
simultaneously.
When doing so, please note:
•
The device synchronizes shock delivery with the next R
wave.
•
Only with Printing option: If the printer is connected to the
device and the
Defibrillation report
menu item is
activated
(see “13.9 Printer settings (only with Printing
, the printer prints a defibrillation
(see “6.16.4 Printing a defibrillation report”,
•
After shock delivery, cardioversion (
SYNC
marking)
remains activated in manual mode. You can set in the
operator menu whether further cardioversion or
defibrillation is to follow cardioversion
mode settings (only with Manual defibrillation option)”,
page 321)
Result
The patient has undergone cardioversion.