MEDUCORE Standard
2
EN
133
6 Operation
Result
The device carries out a cardiac rhythm analysis. The cardiac
rhythm analysis has one of two results:
•
Shock required
(see “ Shock required”, page 133)
or
•
Shock not possible
(see “ Shock not possible”, page 135)
Shock required
The device carries out a cardiac rhythm analysis, charges for shock
delivery and outputs the message:
If, based on the cardiac rhythm analysis, the device determines that
a shock is required, the device outputs the message:
The shock button
flashes and an audible alarm sounds.
Voice prompt
AED instruction
Stand clear of the patient
Stand clear of the patient
Cardiac rhythm is being analyzed
Analysis
Voice prompt
AED instruction
Shock required
Shock required
Press shock button
Press shock button
Risk of injury from electric shock!
The electric shock administered to the patient may injure the user
or bystanders.
Stand clear of the patient.
Keep patient away from liquids (e.g. blood, gel or saline
solution).
Do not touch parts in contact with the patient (e.g. bed frames
or stretchers).
Keep your distance from liquids in contact with the patient.
Clearly warn bystanders to stand clear of the patient or parts in
contact with the patient and to keep their distance from liquids
in contact with the patient.