11 Alarms and faults
* If the device outputs this alarm, it will no longer output a
simultaneously occurring SpO
2
measured value alarm. If signal
quality drops to < 20 %, the device will also stop displaying the
SpO
2
measured value and the plethysmogram.
11.3 Faults
If you are unable to remedy faults with the aid of the table, you
should contact the manufacturer WEINMANN Emergency, or
technicians expressly authorized by WEINMANN Emergency, to
have the device repaired. To avoid more serious damage, do not
continue using the device.
SD card defective
SD card write/read error
, and switch device back on
after at least 30 s
.
Resume session and do not record
any more data.
SD card full
SD card full
, and switch device back on
after at least 30 s
.
Erase data on current SD card.
SpO
2
sensor defective
Pulse oximetry sensor connecting
cable defective
Replace pulse oximetry sensor
connecting cable.
Pulse oximetry sensor defective
Replace pulse oximetry sensor.
SpO
2
signal quality
r
*
SpO
2
signal quality < 40 %
Attach pulse oximetry sensor to
the patient correctly
.
If medically indicated, carry out
suitable treatment.
Update operator menu access
code
(Alarm deactivated after 10 s)
Access code for the operator menu
is 000000 (status when device
supplied)
Wait 10 seconds.
Change operator menu access
code
Alarm
Cause
Remedy