9.6.10 Low Vacuum Generated
Probable Cause
Corrective Action
Vacuum Venturi supply pressure or
flow is insufficient.
Vent isolation valve, MV22, is
closed
Verify a Venturi supply of 75-95 psig
(5.2-6.6 barg) at 50-60 slpm (106-127
cfh) is obtainable.
Open fully