The recommendations are briefly defined as follows:
G.821:
Error performance of an international digital connection operating at a bit rate below the primary rate and forming part of an Integrated Service Digital Network (ISDN)
Long term error performance conducted Out of Service (OOS)
Based on measuring bit errors
Evaluation period of 30 days
Since there is no overhead structure at these bit rates, in-service measurements are extremely difficult
G.826:
End-to-end error performance parameters and objectives for international, constant bit rate digital paths and connections
Long term error performance for Out of Service (OOS) and In-Service Measurement (ISM)
Based on measuring bit errors for connections and block errors for paths
Evaluation period of 30 days
G.828 Analysis
: Error performance parameters and objectives for international constant bit rate synchronous digital paths
Deals exclusively with SDH paths
Long term error performance for Out of Service (OOS) and In-Service
Based on measuring block errors
Evaluation period of 30 days
G.829 Analysis:
Error performance events for SDH Regenerator and Multiplexer sections
Deals exclusively with SDH sections
Error event definition
Based on measuring block errors
M.2100:
Performance limits for bringing into service and maintenance of international multi-operator PDH paths and connections
Deals exclusively with PDH paths, sections and systems
Based on measuring bit errors and block errors
BIS limits for OOS/ISM
Evaluation periods of 15 minutes, 2 hours and 24 hours
First step is a continuity test for 15 minutes
PDH paths are composed of sub-elements of different lengths each with its own set Reference Performance Objectives (RPO)
M.2101 Analysis:
Performance limits and objectives for maintenance and bringing into service SDH paths and multiplex sections.
Deals exclusively with SDH paths and sections
Based on measuring block errors
BIS limits for OOS/ISM
Evaluation periods of 15 minutes, 2 hours, 24 hours and 7 days
First step is a continuity test for 15 minutes
Note:
Only one performance analysis can be performed at a time. To view or enable the M.2100 and M.2101 analyses, the measurement timer has to be set to a determined
period. The table below describes the anomalies evaluated for the performance analysis selected.
7.1.3 General
The General setup page configures the audible alarm and APS protocol settings.
General tab
UX400 10G e-Manual D07-00-058P RevA00
Page 123 of 209
Summary of Contents for UX400
Page 1: ...UX400 10G e Manual D07 00 058P RevA00 Page 1 of 209 ...
Page 33: ...Go back to top Go back to TOC UX400 10G e Manual D07 00 058P RevA00 Page 33 of 209 ...
Page 37: ...Go back to top Go back to TOC UX400 10G e Manual D07 00 058P RevA00 Page 37 of 209 ...
Page 142: ...Go back to top Go back to TOC UX400 10G e Manual D07 00 058P RevA00 Page 142 of 209 ...
Page 189: ...Go back to top Go back to TOC UX400 10G e Manual D07 00 058P RevA00 Page 189 of 209 ...
Page 205: ...Go back to top Go back to TOC UX400 10G e Manual D07 00 058P RevA00 Page 205 of 209 ...