4.11 Additional functions
4-30
4.11.2 Oxygenation
Activation of the function is available through [Menu]
→
[Additional functions]
→
[FiO
2
100% 2
min] or
icon on the main screen.
In the oxygenation mode 100 % oxygen is provided for 2 minutes. Field of oxygen concentration
setting in supplied breathing mixture becomes yellow and contains the countdown timer that
shows the time left before the end of oxygenation. At clicking on this field or icon
oxygena-
tion is disabled.
4.11.3 Suction
Activation of the function is available through [Menu]
→
[Additional functions]
→
[Suction] or
icon
After activation of the function for 3 minutes "Oxygenation" is performed (see it.4.11.2). At this
time the ventilator waits for disconnecting of the patient. If the disconnection does not occur, the
device returns to the previous mode, the "Suction" function is switched off.
If disconnection occurs, the device stops ventilation. Window with the message of disconnection
and with timer showing time of its beginning appears. Operation with the interface and alarms
are blocked. The ventilator is waiting for connection of patient after suction. If connection does
not occur within 2 minutes, the device activates sound alarm.
After connection device resumes ventilation. Oxygenation is swiching for 2 minutes.
Suction can be stopped during operation by disabling "Suction" or "Oxygenation" function
through the menu or through the icon.
Summary of Contents for MV200
Page 2: ......
Page 8: ......
Page 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Page 124: ...4 21 Open valve function 4 76 For your notes ...
Page 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Page 160: ...7 Troubleshooting 7 2 For your notes ...
Page 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Page 166: ...13 commission date mark 12 2 For your notes ...
Page 168: ...14 Maintenance and repair data 14 2 For your notes ...
Page 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Page 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Page 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Page 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...