4.14 Alarm settings menu
4-44
4.14 Alarm settings menu
Enter the window through the main menu of ventilator (see it. 4.6).
Figure 4.9. Alarm threshold parameters
Line of the
menu
Designation, description of parameters
Possible values Default value
FiO
2
FiO
2
sensor
– display of FiO
2
readings
(dashes «---» are displayed in the «
off
» position)
on/off
on
FiO
2
alarms
–
enable/disable FiO
2
alarms
on/off
on
О
2
deviation
– level deviation of measured
concentration from set that triggers the alarm.
1…50 %,
off
5 %
О
2
pressure
(min)
Low oxygen pressure alarm threshold in the input
line. The recommended values are 1.0, 1.5 atm
(kgf/cm
2
). Set "0.0" value only in case of
ventilation from low pressure
О
2
source or air.
0.0, 0.5, 1.0, 1.5
atm
1.0 atm
Low PIP
Low inspiratory pressure alarm threshold. "0"
value means disabled alarm.
0…20 cmH
2
O
0 cmH
2
O
Low PEEP
Low PEEP alarm threshold. "0" value means
disabled alarm.
0…10 cmH
2
O
0 cmH
2
O
MV
MV_max
–
top alarm threshold of minute breathing
volume.
MV_min…60 lpm
12 lpm
MV_min
–
bottom alarm threshold of minute
breathing volume.
0…MV_max
lpm
2 lpm
Vexp
Vexp_max
– top alarm threshold of expiratory
volume.
Vexp_min… 6000
ml
6000 ml
Vexp_min
– bottom alarm threshold of expiratory
volume.
0 … Vexp_max
ml
0 ml
RB
RB_max
– top alarm threshold of respiratory rate.
RB_min…120
1
/min
40
1
/min
RB_min
– bottom alarm threshold of respiratory
rate.
1… RB_max
1
/min
8
1
/min
Capnometer
EtCO
2
_max
– top alarm threshold of
СО
2
partial
pressure in exhaled air [mm Hg].
EtCO
2
_min…100
mmHg
40
EtCO
2
_min
– bottom alarm threshold of
СО
2
partial pressure in exhaled air [mm Hg].
15…
EtCO
2
_max
mmHg
15
Summary of Contents for MV200
Page 2: ......
Page 8: ......
Page 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Page 124: ...4 21 Open valve function 4 76 For your notes ...
Page 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Page 160: ...7 Troubleshooting 7 2 For your notes ...
Page 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Page 166: ...13 commission date mark 12 2 For your notes ...
Page 168: ...14 Maintenance and repair data 14 2 For your notes ...
Page 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Page 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Page 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Page 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...