6.1 Maintenance schedule
6-1
6
MAINTENANCE OF THE DEVICE
Before performing any maintenance, ensure that the device and its accesso-
ries are properly disinfected.
To ensure the proper functioning of the device during its service life make regular maintenance
in accordance with it. 6.1
Maintenance of device is the responsibility of the user of the device and is not included in the
warranty obligations of the manufacturer or supplier.
Maintenance routines do not involve disassembly / assembly of the electronic unit and does not
require specialized skills and knowledge.
If any defects that require disassembly of the electronic unit are found, the device must be
transferred to the repair organizations authorized by Triton Electronic Systems Ltd. that have
the proper qualifications and the necessary equipment.
Maintenance should be carried out with the frequency recommended in it.6.1. Even if the device
is taken out of service and is temporarily stored, at least every 6 months check the operation of
the device and the FiO
2
sensor, perform the training of the built-in rechargeable battery.
6.1 Maintenance schedule
№
Routine
Periodicity
Description
1. Visual inspection and test of controls
monthly
it.6.2
2. Cleaning of fan filters
monthly
it.6.3
3. Check of sealing gaskets of oxygen hose and
water trap
1 time in 3 months
it.6.4
4. Check of the filter-regulator
1 time in 3 months
it.6.5
5. Check of the expiration valve
monthly
it.6.6
6. Check and calibration of the FiO
2
oxygen sen-
sors and the metabolism measurement module
monthly
it.6.7
7. Check of built-in accumulator
1 time in 3 months
it.6.8
8. Calibration of the exhalation flow sensor
monthly
it.4.16.3
9. Check of the compliance of the set and dis-
played volume of inspiration and expiration
monthly
it.6.9
10. Check of PEEP
monthly
it.6.10
11. Check and calibration of mainstream CO
2
sen-
sor
monthly*
it.6.11
12.
Breathing circuit calibration
monthly**
it.4.16.5
* - Calibration of mainstream CO
2
sensor is performed with every replacement of the airway adapter or
breathing circuit
** - Calibration of the breathing circuit is intended to adapt the device to the breathing circuit. It is recom-
mended to perform it at the device turning on and especially at changing of breathing circuit type.
Summary of Contents for MV200
Page 2: ......
Page 8: ......
Page 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Page 124: ...4 21 Open valve function 4 76 For your notes ...
Page 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Page 160: ...7 Troubleshooting 7 2 For your notes ...
Page 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Page 166: ...13 commission date mark 12 2 For your notes ...
Page 168: ...14 Maintenance and repair data 14 2 For your notes ...
Page 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Page 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Page 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Page 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...