4.2 Reserve power supply
4-4
4.2.2 Features of reserve power supply operation
Table 4.1. Reserve power supply conditions
Reserve power supply condition
Light indicator
“BATTERY”
Graphical
symbol on the
display
Alarm priority
Charge from the mains
Green
Green
No
Fully charged
Off
Green
No
Discharge
Yellow
Yellow
Medium*
Fully discharged
Blinking
yellow
Blinking
Red
High**
Malfunction
Red
Blinking
Red
-
High***
* - "No mains voltage".
** - "< 10 min left” or “Accumulator is discharged"
.
*** - “
Accumulator malfunction”
Fully charge built-in accumulator before every using of ventilator with the
connection to the patient.
Presence of the accumulator in the device demands considering features of
its operation (see below).
•
Device operating time in the independent mode depends on the accumulator capacity, its
previous charge level and time passed after that, quality of the accumulator, terms of its
operation, and also from the operating mode of the device.
•
During long-term operation or storage there can be reduction of accumulator capacity,
therefore time of independent operation of the device can be reduced. That can not be a
reason for claims to the manufacturer.
•
For the maintenance of capacity and increasing of the service life of the accumulator it is
necessary to conduct periodically its training (at least once a half a year) (see Section 6).
If the device is connected to the mains power, but the indicator “accumulator” is red, it means
that charging device and/or the accumulator are faulty and need repair. Operation of such
device is forbidden, because it does not provide possibility of operation at interruptions of power
supply.
In some cases, at malfunction of the accumulator, the device can continue ventilation in the
presence of mains supply. Also at the detection of the accumulator failure during the device
operation, ventilation does not stop, but the accumulator indicator on the screen blink with red
color.
In this situation it is necessary to prepare immediately other means for the
patient ventilation and be ready to use them. When operating with the faulty
accumulator, power supply interruptions can cause switching off the device
and the ventilation termination without any warnings.
Responsibility for negative consequences of operation of the device with
the faulty accumulator lies on medical staff of the medical facility.
Summary of Contents for MV200
Page 2: ......
Page 8: ......
Page 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Page 124: ...4 21 Open valve function 4 76 For your notes ...
Page 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Page 160: ...7 Troubleshooting 7 2 For your notes ...
Page 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Page 166: ...13 commission date mark 12 2 For your notes ...
Page 168: ...14 Maintenance and repair data 14 2 For your notes ...
Page 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Page 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Page 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Page 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...