3.3 Preparing for operation
3-10
Figure 3.10 – Assembly of the breathing circuit with the mask for NIV (optional)
Figure 3.11 – Removing of the liquid from the removable water trap chamber
For connection of the masks manufactured by BMC Medical or Intersurgical
with breathing circuit use connector 22F-22F REF1967 Intersurgical Ltd, UK.
Breathing circuit should contain
virus-bacterial
filter.
Breathing circuit should contain at least one water trap in the expiration
line. Otherwise the manufacturer is not liable for the consequences of the
device malfunctioning.
When using heat-moisture exchange filters it is necessary to consider the
features of their operation (see it. 4.19).
During the ventilation it is necessary to monitor the level of liquid in the wa-
ter trap chamber at the bottom part of the expiration valve and if necessary
to empty it as shown in Figure 3.11.
Summary of Contents for MV200
Page 2: ......
Page 8: ......
Page 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Page 124: ...4 21 Open valve function 4 76 For your notes ...
Page 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Page 160: ...7 Troubleshooting 7 2 For your notes ...
Page 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Page 166: ...13 commission date mark 12 2 For your notes ...
Page 168: ...14 Maintenance and repair data 14 2 For your notes ...
Page 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Page 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Page 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Page 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...