4.12 Display settings menu
4-38
•
current loop becomes the reference at the end of the respiratory cycle;
•
reference loop is fixed in grey color;
•
line of reference loop fixing time is displayed.
At the repeated pressing of REF, the previous loop is selected as reference. The reference loop
is updated at every new cycle. Time line will be hidden.
In the mode of displaying of 4 curves the last selected reference loop is displayed.
Displaying of iSV graph
iSV graph allows to estimate the optimal choice of parameters of ventilation by algorithms of
current ventilation mode. The iSV graph displays the respiratory rate and tidal volume scales.
iSV graph shows:
•
limits of safe ventilation by RB and Vt, top and bottom values (in graphical and numerical
view);
•
target values of RB and Vt (in graphical and numerical view);
•
real values of RB and Vt (in graphical and numerical view);
•
target minute ventilation values in % and ml/min;
•
adaptation minute ventilation values in % and ml/min;
•
values of the current inspiratory pressure for hardware breaths;
•
values of the current support pressure for spontaneous breaths.
Example of realization of graph:
Inspiraiton volume Vtmax can be limited in [Menu]
→
[Ventilation parameters]
→
[iSV
parameters].
Changing of the graph indication mode does not effect on the current lung
ventilation mode.
Summary of Contents for MV200
Page 2: ......
Page 8: ......
Page 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Page 124: ...4 21 Open valve function 4 76 For your notes ...
Page 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Page 160: ...7 Troubleshooting 7 2 For your notes ...
Page 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Page 166: ...13 commission date mark 12 2 For your notes ...
Page 168: ...14 Maintenance and repair data 14 2 For your notes ...
Page 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Page 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Page 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Page 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...