3.3 Preparing for operation
3-11
Figure 3.12 – Connection of the coaxial breathing circuit
3.3.5 Connecting to the oxygen source
The device can be connected to oxygen sources of various types (main gas supply, cylinders,
oxygen concentrators). For connection to oxygen sources the low pressure oxygen hose comply
with EN ISO 5359 is used. For connection make the follows:
•
Connect the filter-regulator from the delivery kit to the “
О
2
” inlet on the rear panel of the elec-
tronic unit using the cap nut with the seal.
•
Connect the supplied oxygen hose with NIST connector to the filter-regulator on the rear
panel of device (see Figure 3.13). Thoroughly insert the oxygen hose nut into filter-regulator
threaded inlet and turn as shown on the Figure 3.13.
Using of the coaxial (or dual chamber) breathing circuit is recommended for
short-term ventilation without humidifier and with HME filter.
When using the humidifier because of design features of such circuits gra-
dual accumulation of moisture and occlusion of the breathing circuit can
occur.
Summary of Contents for MV200
Page 2: ......
Page 8: ......
Page 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Page 124: ...4 21 Open valve function 4 76 For your notes ...
Page 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Page 160: ...7 Troubleshooting 7 2 For your notes ...
Page 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Page 166: ...13 commission date mark 12 2 For your notes ...
Page 168: ...14 Maintenance and repair data 14 2 For your notes ...
Page 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Page 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Page 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Page 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...