4.16 Service menu
4-54
No
Message
Alarm condition
Operator’s actions
12
Incorrect coeff. for differential
press. sensor
Ventilation controller malfunc-
tion
Contact the customer service
13
Error: no response from power
unit
Power unit malfunction
Contact the customer service
14
Error: temperature sensor
error in power unit
Power unit malfunction
Contact the customer service
15
Error: low ambient air
temperature
Power unit malfunction
Contact the customer service
16
Calibration task stop
Calibration task failure
Repeat the calibration. If mes-
sage appears again contact the
customer service
17
Monotony error in input table
User incorrectly followed the
calibration instructions on the
screen or malfunction of the
device
Repeat the calibration. If
message appears again contact
the customer service
18
Coefficients writing error
Ventilation controller malfunc-
tion
Contact the customer service
19
Incorrect coeff.for
press.sensor in exp.line
Ventilation controller malfunc-
tion
Contact the customer service
4.16.4 Selection of flow correction conditions on the inspiration and the expiration
On the inspiration the device uses the flow sensors, operating on the principle of measuring the
mass of the passed gas. On the expiratory flow sensor measures the volume of the passed
breathing mixture. It is possible to use different types of breathing gas humidification in the
device.
“Insp. conditions” and “Exp. conditions” parameters are intended for selection of conditions for
the conversion of the breathing mixture mass passed through the device to the appropriate
volume. Weight and volume are connected by the equation that contains the partial pressures of
all gas fractions, temperature and the absolute pressure of the mixture.
Select the parameters through the menu: [Menu]
→
[Service menu]
→
[Calibration]
→
[CIVL]
→
[Insp. (Exp.) conditions].
The device provides 4 types of inspiration and expiratory flow correction:
•
ATP (Ambient Temperature, Pressure) - the conversion is carried out for the 30 % relative
humidity and ambient temperature of 25 °C.
•
ATPS (Ambient Temperature, Pressure, Saturated – 100 % humidity), this correction is
similar to ATP at 100 % relative humidity.
•
ATPD (Ambient Temperature, Pressure, Dried – 0 % humidity), this correction is similar to
ATP at 0 % relative humidity.
•
BTPS (Body Temperature, Pressure, Saturated – ambient pressure, 100 % relative
humidity).
When working with the bag select ATP correction, when using the humidifier select BTPS.
Incorrect type of correction would lead to the increasing of measurement errors of inspiration
and expiration volume.
Summary of Contents for MV200
Page 2: ......
Page 8: ......
Page 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Page 124: ...4 21 Open valve function 4 76 For your notes ...
Page 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Page 160: ...7 Troubleshooting 7 2 For your notes ...
Page 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Page 166: ...13 commission date mark 12 2 For your notes ...
Page 168: ...14 Maintenance and repair data 14 2 For your notes ...
Page 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Page 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Page 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Page 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...