4.16 Service menu
4-56
No
Message
Alarm condition
Operator’s actions
10
Could not switch CIVL to
calibration mode
Ventilation controller does
not respond to the requests
of breathing circuit
calibration.
Contact the customer service.
11
Monotony error in input table User incorrectly followed the
breathing circuit calibration
instructions on the screen or
malfunction of the device.
Repeat the breathing circuit
calibration. If message appears
again contact the customer
service.
12
Coefficients writing error
Ventilation controller
malfunction.
Contact the customer service.
13
Breathing circuit calibration
time out
No
messages
from
ventilation controller about
results of breathing circuit
calibration.
Contact the customer service.
14
Gas mixer malfunction
Gas mixer malfunction.
Contact the customer service.
15
Power unit does not respond Power unit does not respond
to the requests of ventilation
controller.
Contact the customer service.
Summary of Contents for MV200
Page 2: ......
Page 8: ......
Page 25: ...2 3 Front panel 2 9 2 3 Front panel Figure 2 1 Front view of the electronic unit ...
Page 124: ...4 21 Open valve function 4 76 For your notes ...
Page 158: ...6 11 Check and calibration of mainstream CO2 sensor 6 8 For your notes ...
Page 160: ...7 Troubleshooting 7 2 For your notes ...
Page 162: ...8 Transportation 9 Storage 10 Disposal 10 2 For your notes ...
Page 166: ...13 commission date mark 12 2 For your notes ...
Page 168: ...14 Maintenance and repair data 14 2 For your notes ...
Page 196: ...Appendix 2 9 Ideal body weight of the patient IBW App 2 10 For your notes ...
Page 200: ...Appendix 3 Electromagnetic environment App 3 4 For your notes ...
Page 201: ...Appendix 4 Pneumatic scheme of ventilator App 4 1 APPENDIX 4 PNEUMATIC SCHEME OF VENTILATOR ...
Page 202: ...Appendix 4 Pneumatic scheme of ventilator App 4 2 For your notes ...